Por favor, use este identificador para citar o enlazar este ítem: https://doi.org/10.1007/s10661-016-5674-8

Título: Study of the effect of sample pressure on in situ BTEX chromatographs
Fecha de publicación: 12-nov-2016
Editorial: Springer
Cita bibliográfica: Environmental Monitoring and Assessment, 2016, Vol. 188 665
ISSN: Print: 0167-6369
Electronic: 1573-2959
Palabras clave: Air monitoring
Data quality
Sample pressure
Benzene
Chromatography
Resumen: In this paper, the influence of sample pressure on benzene measurements obtained with two automated in situ gas chromatographs is studied. The analysers were calibrated using a non-linear regression at 293 ± 1 K and 101.3 ± 0.2 kPa. A gas mixture of benzene in air (5 μg/m3) was produced and measured at calibration conditions. Subsequently, the sample pressure was changed to 80 and 110 kPa. Differences in readings were observed even though the pressure compensators were on, indicating incorrect performance of this tool. Further tests with two different benzene in air mixtures (5 and 40 μg/m3) at 80, 90, 105 and 110 kPa were also carried out. Results showed that the analysers take air from one or several unidentified inlets (apart from the sampling port) when the sampling pressure is lower than the atmospheric one. This is usually the case in air monitoring stations, so this phenomenon is particularly important as systematic biases could be affecting air quality data.
Autor/es principal/es: Romero Trigueros, Cristina
Doval Miñarro, Marta
González Duperón, Esther
Marzal Martínez, Francisco José
González Ferradás, Enrique
Versión del editor: https://link.springer.com/article/10.1007/s10661-016-5674-8
URI: http://hdl.handle.net/10201/148458
DOI: https://doi.org/10.1007/s10661-016-5674-8
Tipo de documento: info:eu-repo/semantics/article
Número páginas / Extensión: 8
Derechos: info:eu-repo/semantics/openAccess
Attribution-NonCommercial-NoDerivatives 4.0 Internacional
Descripción: ©2016. This manuscript version is made available under the CC-BY-NC-ND 4.0 license http://creativecommons.org/licenses/by-nc-nd/4.0/ This document is the accepted version of a Published Work that appeared in final form in Environmental Monitoring and Assessment. To access the final edited and published work see https://doi.org/10.1007/s10661-016-5674-8
Aparece en las colecciones:Artículos

Ficheros en este ítem:
Fichero Descripción TamañoFormato 
Romero-Trigueros et al., 2016. Study of the effect of sample pressure on in situ BTEX chromatographs.doc253 kBMicrosoft WordVisualizar/Abrir


Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons