Palabras clave: | Range of motion Upper limb joints Biomedical technology |
Resumen: | The aim of this study was to evaluate the validity of the Veloflex infrared dynamic angle-meter (Veloflex-IDA) and the intra- and inter-rater reliability when measuring the ranges of motion (ROMs) of the upper limb joints. Thirty-five healthy and 20 symptomatic participants were evaluated. Twelve upper limb movements were measured in two sessions with the Veloflex-IDA, which is a device composed of a camera that tracks the trajectory of retro-reflective markers. In addition, a goniometer was used in the first session to evaluate concurrent validity. Validity and agreement were evaluated by Pearson correlation coefficient (r) and Bland-Altmann plots. Intra- and inter-rater reliability were evaluated using intra-class correlation (ICC), standard error of measurement (SEM), and minimal detectable change (MDC). Both instruments showed excellent correlation for all movements (r range from 0.992 to 0.999). The intra- and inter-rater reliability were excellent (ICC range from 0.95 to 0.99 and 0.90 to 0.98, respectively). Intra-rater reliability showed SEMs <1.38% and <5.19% and inter-rater reliability SEMs <2.26% and <5.22% for asymptomatic and symptomatic, respectively. Veloflex-IDA is a valid and reliable alternative to measure the upper limb joints' ROM and it can be used in clinical practice and research after basic training. |
Autor/es principal/es: | Martín San Agustín, Rodrigo García Vidal, José A. Cánovas Ambit, Germán Arenas Dalla Vecchia, Aurelio López Nicolás, Manuel Medina Mirapeix, Francesc |
Facultad/Departamentos/Servicios: | Facultad de Química, Departamento de Electromagnetismo y Electrónica |
URI: | http://hdl.handle.net/10201/138401 |
DOI: | https://doi.org/10.3390/jcm8111851 |
Tipo de documento: | info:eu-repo/semantics/article |
Número páginas / Extensión: | 11 |
Derechos: | info:eu-repo/semantics/openAccess Attribution-NonCommercial-NoDerivatives 4.0 Internacional |
Descripción: | ©2019. This manuscript version is made available under the CC-BY 4.0 license http://creativecommons.org/licenses/by-nc-nd/4.0/ This document is the Published, version of a Published Work that appeared in final form in Journal of Clinical Medicine . To access the final edited and published work see https://doi.org/10.3390/jcm8111851 |
Matería temporal: | Siglo XXI |
Matería geográfica: | Laboratorio clínico |
Aparece en las colecciones: | Artículos: Electromagnetismo y Electrónica |
Ficheros en este ítem:
Fichero | Descripción | Tamaño | Formato | |
---|---|---|---|---|
Validation of Veloflex.pdf | Validation of an optoelectronic system for measuring range of motion | 1,44 MB | Adobe PDF | Visualizar/Abrir |
Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons