Por favor, use este identificador para citar o enlazar este ítem: https://doi.org/10.3390/jcm8111851

Título: Validity and realiability of a new optoelectronic system for measuring active range of motion of upper limb joints in asymptomatic and symptomatic subjects
Fecha de publicación: 2-nov-2019
Fecha de defensa / creación: 9-oct-2019
Editorial: MDPI
Cita bibliográfica: Journal of Clinical Medicine, 2019, 8, 1851
ISSN: 2077-0383
Materias relacionadas: CDU::6 - Ciencias aplicadas::62 - Ingeniería. Tecnología
Palabras clave: Range of motion
Upper limb joints
Biomedical technology
Resumen: The aim of this study was to evaluate the validity of the Veloflex infrared dynamic angle-meter (Veloflex-IDA) and the intra- and inter-rater reliability when measuring the ranges of motion (ROMs) of the upper limb joints. Thirty-five healthy and 20 symptomatic participants were evaluated. Twelve upper limb movements were measured in two sessions with the Veloflex-IDA, which is a device composed of a camera that tracks the trajectory of retro-reflective markers. In addition, a goniometer was used in the first session to evaluate concurrent validity. Validity and agreement were evaluated by Pearson correlation coefficient (r) and Bland-Altmann plots. Intra- and inter-rater reliability were evaluated using intra-class correlation (ICC), standard error of measurement (SEM), and minimal detectable change (MDC). Both instruments showed excellent correlation for all movements (r range from 0.992 to 0.999). The intra- and inter-rater reliability were excellent (ICC range from 0.95 to 0.99 and 0.90 to 0.98, respectively). Intra-rater reliability showed SEMs <1.38% and <5.19% and inter-rater reliability SEMs <2.26% and <5.22% for asymptomatic and symptomatic, respectively. Veloflex-IDA is a valid and reliable alternative to measure the upper limb joints' ROM and it can be used in clinical practice and research after basic training.
Autor/es principal/es: Martín San Agustín, Rodrigo
García Vidal, José A.
Cánovas Ambit, Germán
Arenas Dalla Vecchia, Aurelio
López Nicolás, Manuel
Medina Mirapeix, Francesc
Facultad/Departamentos/Servicios: Facultad de Química, Departamento de Electromagnetismo y Electrónica
URI: http://hdl.handle.net/10201/138401
DOI: https://doi.org/10.3390/jcm8111851
Tipo de documento: info:eu-repo/semantics/article
Número páginas / Extensión: 11
Derechos: info:eu-repo/semantics/openAccess
Attribution-NonCommercial-NoDerivatives 4.0 Internacional
Descripción: ©2019. This manuscript version is made available under the CC-BY 4.0 license http://creativecommons.org/licenses/by-nc-nd/4.0/ This document is the Published, version of a Published Work that appeared in final form in Journal of Clinical Medicine . To access the final edited and published work see https://doi.org/10.3390/jcm8111851
Matería temporal: Siglo XXI
Matería geográfica: Laboratorio clínico
Aparece en las colecciones:Artículos: Electromagnetismo y Electrónica

Ficheros en este ítem:
Fichero Descripción TamañoFormato 
Validation of Veloflex.pdfValidation of an optoelectronic system for measuring range of motion1,44 MBAdobe PDFVista previa
Visualizar/Abrir


Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons