Por favor, use este identificador para citar o enlazar este ítem: https://doi.org/10.14670/HH-18-342

Título: Osseointegration of a novel 3D porous Ti-6Al-4V implant material – Histomorphometric analysis in rabbits
Fecha de publicación: 2021
Editorial: Universidad de Murcia, Departamento de Biologia Celular e Histiologia
Cita bibliográfica: Histology and Histopathology Vol. 36, nº8 (2021)
ISSN: 0213-3911
1699-5848
Materias relacionadas: CDU::6 - Ciencias aplicadas::61 - Medicina::616 - Patología. Medicina clínica. Oncología
Palabras clave: Porous Titanium Implants
Osseohealing
Osseointegration
Bone Remodeling
Histology
BoneImplant Contact (BIC)
Resumen: Porous structure properties are known to conduct initial and long-term stability of titanium alloy implants. This study aims to assess the histomorphometric effect of a 3-D porosity in Ti-6Al-4V implants (PI) on osseointegration in comparison to solid Ti-6Al-4V implants (SI). The PI was produced in a spaceholder method and sintering and has a pore size of mean 400 µm (50 µm to 500 µm) and mimics human trabecular bone. Pairs of PI and equal sized SI as reference were bilaterally implanted at random in the lateral femoral condyle of 16 Chinchilla-Bastard rabbits. The animals were sacrificed after 4 and 12 weeks for histomorphometric analysis. The histomorphometric evaluation confirmed a successful short-term osseohealing (4 weeks) and mid-term osseoremodeling (12 weeks) for both types of implants. The total newly formed bone area was larger for PI than for SI after 4 and 12 weeks, with the intraporous bone area being accountable for the significant difference (p<0.05). A more detailed observation of bone area distribution revealed a bony accumulation in a radius of ±500 µm around the implant surface after remodeling. The boneto-implant contact (BIC) increased significantly (p<0.05) from 4 to 12 weeks (PI 26.23% to 42.68%; SI 28.44% to 47.47%) for both types of implants. Due to different surface properties, however, PI had a significant (p<0.05) larger absolute osseous contact (mm) to the implant circumference compared to the SI (4 weeks: 7.46 mm vs 5.72 mm; 12 weeks: 11.57 mm vs 9.52 mm [PI vs. SI]). The regional influences (trabecular vs. cortical) on bone formation and the intraporous distribution were also presented. Conclusively, the porous structure and surface properties of PI enable a successful and regular osseointegration and enhance the bony fixation compared to solid implants under experimental conditions.
Autor/es principal/es: Frosch, Stephan
Buchhorn, Gottfried
Krohn, Sebastian
Lehmann, Wolfgang
Frosch, Karl-Heinz
Füzesi, László
Frosch, Alice
URI: http://hdl.handle.net/10201/127744
DOI: https://doi.org/10.14670/HH-18-342
Tipo de documento: info:eu-repo/semantics/article
Número páginas / Extensión: 10
Derechos: info:eu-repo/semantics/openAccess
Attribution-NonCommercial-NoDerivatives 4.0 Internacional
Aparece en las colecciones:Vol.36, nº8 (2021)

Ficheros en este ítem:
Fichero Descripción TamañoFormato 
Frosch-36-879-888-2021.pdf3,8 MBAdobe PDFVista previa
Visualizar/Abrir


Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons